How to use SNTP on DL850 ScopeCorder?

​How to use SNTP on DL850 ScopeCorder

The sequence below is very important for SNTP to work properly.

1. Set DATE/TIME and GMT Time Difference to your location FIRST. EDT is -4 hours, EDT is -5 hours - as examples.

2. Set a SNTP Server for GMT in SNTP settings (64.90.182.55 in New York City for example - this is GMT, not EDT or EST);

3. Set up TCP IP as usual.

4. Under SNTP function, press ADJUST .. this must be performed as the last step.

Related Products & Solutions

DLM2000 Mixed Signal Oscilloscopes

200, 350, and 500MHz mixed-signal oscilloscopes for every engineer. Best-in-class performance in usability, acquisition, analysis, and display—all at a price you can digest. Options include serial bus, vehicle bus, and power supply analysis functions.

Data Acquisition (DAQ)

Simplify complex testing with DAQ systems that flex to your needs. Capture and analyze any physical or electrical signal with ease, transforming signals into actionable insights that speed innovation and improve reliability.

High Speed Data Acquisition

  • PC-based, streaming, local or remote operation
  • 20+ modules, isolated and versatile inputs
  • Up to 200 MS/s or 640 ch

Isolated Oscilloscopes | ScopeCorders

  • An integrated measurement system for every electromechanical application
  • Modular platform combines oscilloscope and DAQ functionality 
  • Capture high-speed transients and low-speed trends 

Mixed Signal Oscilloscopes

  • Analyze analog and digital signals simultaneously
  • Advanced triggering and high-speed waveform capture
  • Power analysis, serial bus analysis, & switching loss

Oscilloscopes

Accelerate design and troubleshooting with oscilloscopes that combine high-speed sampling, long record lengths, and mixed-signal analysis. From quick debug to advanced R&D, Yokogawa oscilloscopes help you see deeper, work faster, and design smarter.

Precision Making

Top