701267 Analog Voltage Input Module (DISCONTINUED)

Notice: This product was discontinued on Sep 29, 2017. See this replacement product:

High-Voltage 100 kS/s sample rate, 16-Bit resolution, 40 kHz bandwidth, 2 channels, Isolated, 850 V*3 Maximum input voltage, ±0.25% DC accuracy, with RMS, and high noise immunity.

ITEM

SPECIFICATIONS

Input channels

2

Input type

Isolated unbalanced

Input coupling

AC, DC, GND, AC-RMS, and DC-RMS

Input connector

Plug-in terminal (safety terminal)

Input impedance

1 MΩ±1%, approx. 35 pF

Maximum sample rate

100 kS/s

Frequency range

(-3dB)*1

Waveform observation mode

DC to 40 kHz

RMS observation mode

DC, 40 Hz to 10 kHz

A/D conversion resolution

16-bit (2400 LSB/div)

Voltage-axis sensitivity setting

20 mV/div to 200 V/div (1-2-5 steps)

Maximum input voltage (1 kHz or less)

 

In combination with 758933+701954 (1:1)

850 V (DC+ACpeak)

Direct input

850 V (DC+ACpeak)

Maximum allowable common mode voltage (1 kHz or less)

 

In combination with 758933+701954 (1:1)*9

H side: 700 Vrms (CAT @), L side: 400 Vrms (CAT @)

Direct input H/L sides:

30 Vrms (42 VDC+ACpeak)*11

Vertical (voltage) axis accuracy*1

 

Waveform observation mode

DC accuracy ±(0.25% of 10 div)

RMS observation mode

DC accuracy ±(1.0% of 10 div)

AC accuracy (sinewave input)

±(1.5% of 10 div) At frequency of 40 Hz to 1 kHz

AC accuracy (crest factor 2 or less)

±(2.0% of 10 div) At frequency of 40 Hz to 1 kHz

AC accuracy (crest factor 3 or less)

±(3.0% of 10 div) At frequency of 40 Hz to 1 kHz

Temperature coefficient (Waveform observation mode)

 

Zero point

±(0.02% of 10 div)/: (Typ.)

Gain

±(0.02% of 10 div)/: (Typ.)

Bandwidth limit

OFF/100 Hz/1 kHz/10 kHz

Response time (RMS observation mode)

 

Rising

(0 to 90% of 10 div) 100 ms (Typ.)

Falling

(100 to 10% of 10 div) 250 ms (Typ.)

Weight

Approx. 250g

Looking for more information on our people, technology and solutions?


연락처

Precision Making

Top