The AQ23211A is a single-channel optical power meter module for the AQ2300 Series Test Platform. Its wide measurement range and fast averaging capability support optical device evaluation, optical transceiver testing, and loss measurements of passive optical components.
The module also provides an analog output for monitoring optical power changes with external instruments. It can be combined with AQ2300 Series light sources, source measure units, optical attenuators, and optical switches to build flexible manual or automated test systems.
By synchronizing the source measure unit and optical power meter and utilizing the sweep function, the I-L characteristics of LD
modules can be measured.
[Measurement items] I-L measurement, DC extinction ratio, PD current, Modulator current

Optical transceivers’ optical power can be measured.
[Measurement items] Optical power, Optical input interruption alarm

Optical insertion loss of optical fibers, etc. can be measured.
[Measurement items] Optical insertion loss

| Items | Specifications | |
|---|---|---|
| Model | AQ23211A | |
| Number of channels (Slot widths) | 1 channel (1 slot) | |
| Detector type | InGaAs | |
| Wavelength range | 800 to 1700 nm | |
| Power range (CW) | −90 to +15 dBm*1 | |
| Minimum display resolution | 1/10000 | |
| Applicable fiber | ≤62.5/125 µm (GI), NA ≤0.275; SMF (ITU-T G.652.D)*3 | |
| Uncertainty under reference conditions | ±2.5%*2, *3 | |
| Total uncertainty | ±5% ±5 pW*2, *4 | |
| Polarization dependence | 0.02 dBp-p (typ.)*2, *5 | |
| Linearity | ±0.02 dB ±5 pW*2, *6 | |
| Noise level | 5 pW or less*7 | |
| Averaging time (minimum) | 20 µs | |
| Analog output | Mode | AUTO, LINEAR, LOG |
| Output voltage | Approx. 0 to 2 V / Approx. 0 to 5 V | |
| Connector type | BNC connector | |
| Output impedance | Approx. 100 Ω | |
| Optical connector | AQ9335C connector adapter*14 | |
| Calibration cycle | 1 year | |
| Dimensions (excluding protrusions) | 106.5 (H) × 31 (W) × 321.5 (D) mm | |
| Weight | Approx. 0.6 kg | |