720254 4-Channel High-Resolution Isolation Module

The Yokogawa 720254-4 is a 4-channel High-Resolution Isolation Module that is capable of 300 kHz of bandwidth at an ultra high 16-bit resolution and ±0.25% accuracy.

With a sampling rate of 1 MS/s, high noise immunity, low noise, and the ability to deliver 600 V isolated measurements (powered by Yokogawa’s IsoPRO™ isolation technology), the 720254-4 combines accuracy, reliability, and safety in one compact design.

This module is perfect for long-term monitoring for power generators and multi-output power supplies. It also excels in dynamic analysis of multi-channel data recording applications.

Key Features

  • Ultra High Resolution: 1 MS/s sampling with 16-bit resolution delivers accurate waveform capture for high-frequency, high-voltage applications

  • Optically Isolated Channels: Each input channel is optically isolated, providing robust protection against ground loops and common-mode interference while maintaining accurate differential measurements even in high-voltage systems. Using high-speed optical fiber-based transmission, IsoPROTM technology enables 600 V isolated measurements*.

  • Consistent Signal Integrity: Maintains low noise and high accuracy across a wide measurement bandwidth

  • Seamless System Integration: Fully compatible with Yokogawa’s ScopeCorder (DL950) and High-Speed DAQ (SL2000) platforms, the 720254-4 extends system capability without adding setup complexity

  • Dynamic Analysis/Long Term Monitoring: Perfect for long-term monitoring for power generators and multi-output power supplies, and dynamic analysis of multi-channel data recording applications

* in combination with the 700929, 702902, or 701947 probes

4-CH 1 MS/s, 16-Bit Isolation Module (720254)
Item Specifications
Standard operating conditions Temperature: 23 ±5 °C
Humidity: 20 to 80 % RH
After a 30 minute warm-up and after calibration
Effective measurement range 20 div (display range: 10 div)
Number of input channels 4
Input coupling settings AC, DC, GND
Maximum sample rate 1 MS/s
Input type Isolated unbalanced
Frequency characteristics1 (−3 dB point when a sine wave of amplitude ±3 div is applied) DC to 300 kHz
Voltage scale setting 10 mV/div to 50 V/div (1-2-5 steps) (when using 1:1 probe factor)
Maximum input voltage (at a frequency of 1 kHz or less)
Combined with the 700929 (10:1), 702902 (10:1), or 701947 (100:1)2 600 V (DC+ACpeak)
Combined with the 701901 + 701954 (1:1)4 200 V (DC+ACpeak) (as a value that meets the safety standard)
400 V (DC+ACpeak) (maximum allowable voltage, as a value that does not damage the instrument when applied)
Direct input (cable that does not comply with the safety standards)6 42 V (DC+ACpeak)
Maximum rated voltage to ground – between input and case and between inputs (at a frequency of 1 kHz or less)
Safety standard’s working voltage  
Combined with 700929 (10:1), 702902 (10:1), 701947 (100:1)3 or combined with 701901 + 701954 (1:1)5 300 Vrms CAT II, 400 Vrms (measurement category Other (O), 1500 V transient overvoltage)
Direct input (cable that does not comply with the safety standards)7 42 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical (voltage) axis accuracy DC accuracy1 ±(0.25 % of 10 div)
Input connector BNC connector (isolated type)
Input impedance 1 MΩ ±1 %, approx. 35 pF
−3 dB point when AC coupled 1 Hz or less (0.1 Hz or less when using the 700929 or 702902, 0.01 Hz or less when using the 701947)
Common mode rejection ratio 80 dB (50/60 Hz) or more (typical value8)
Residual noise level (input section shorted) ±0.05 div (typical value8) 10 mV/div, 20 mV/div
±0.025 div (typical value8) 50 mV/div to 20 V/div
Withstand voltage 2300 Vrms for 2 seconds (across each input terminal and earth) (60 Hz)
Insulation resistance 500 VDC, 10 MΩ or more (across each input terminal and earth)
A/D converter resolution 16 bits (2400 LSB/div)
Temperature coefficient Zero point: ±(0.02 % of 10 div)/°C (typical value8)
Gain: ±0.02%/°C (typical value8)
Bandwidth limit Select from Full and 40 kHz
Cutoff characteristics: –12 dB/OCT (typical value8)
Digital filter
  Cutoff frequency: 12.8 kHz, 6.4 kHz, 3.2 kHz, 1.6 kHz, 800 Hz, 400 Hz, 200 Hz, 100 Hz, 50 Hz, 25 Hz, 12.5 Hz, 6.25 Hz
  Filter type: IIR
Probe attenuation settings Voltage probe: 1:1, 2:1, 5:1, 10:1, 20:1, 50:1, 100:1, 200:1, 500:1, 1000:1, 2000:1, 5000:1, 10 k:1, 20 k:1, 50 k:1
Current probe: 0.1 A:1 V, 0.2 A:1 V, 0.5 A:1 V, 1 A:1 V, 2 A:1 V, 5 A:1 V, 10 A:1 V, 20 A:1 V, 50 A:1 V, 100 A:1 V, 200 A:1 V, 250 A:1 V, 400 A:1 V, 500 A:1 V, 1 kA:1 V, 2 kA:1 V, 5 kA:1 V, 10 kA:1 V, 20 kA:1 V, 50 kA:1 V
Wiring / measurement points
Wiring diagram for the numbered measurement points 2 to 7: combined with the 700929, 702902 or 701947 probe; combined with the 701901 + 701954 cable; direct input (cable that does not comply with the safety standards).
 

1. Value measured under standard operating conditions.
8. Typical values represent typical or average values. They are not strictly warranted.