Clamp-on Tester CL130 (ACA, ACV, 600A) (DISCONTINUED)

Notice: This product was discontinued on Oct 31, 2019. See this replacement product:
  • AC current measurement with current ranges up to 600A
  • AC voltage and resistance measurements
  • 30mm in diameter of measurable conductor
  • Mean value display
  • Data hold function
  • Approved for conformity to safety standards EN61010-1,
  • EN61010-2-031, EN61010-2-032 (CAT. Ill 600 V)
Display (LCD, Digital Counts) 1999
AC Current Accuracy (200A Reference) 1.5+6(50/60Hz)
2.0+5(40~1kHz)
AC Current Accuracy (600A Reference) 1.0+3(50/60Hz)
2.0+5(40~1kHz)
AC Voltage Accuracy (200V/600V Reference) 1.0+2(50/60Hz)
1.5+4(40~1kHz)
Resistance Accuracy (200Ω reference) 1.2+4, Beeps at below 30Ω (continuity check)
Method of detection Mean Value
Response Time Approx. 1 seconds (approx. 2 seconds on resistance range)
Range Switching Manual-range
Data hold On all range
Operating temperature and humidity -10-50°C. (no condensation)
up to 30°C, 90% RH
up to 40°C, 75% RH
up to 50°C, 45% RH
Influence of external magnetic field 2A or less at 400A/m
Influence of conductor position ± 2% or less
Safety Standard Conforms EN 61010-1, EN61010-2-031, EN61010-2-032
Circuit voltage 600Vrms or less
Withstanding voltage 5.55kV AC for one minute
Power Supply 6F22(006P)9Vx1 or 6LR61x2
Battery Life1 200 hours
Consumed current Approx. 2mA
Auto power-off Approx. 10 minutes
Diameter of measurable conductor 30mm at maximum
Dimensions Approx. 93(W)x210(H)x40(D)mm
Weight (Approx.) 400g
Accessories Clamp Adapter (Model 99025), Carrying Case (Model 93032), Test Lead (Model 98010), User's manual (IM CL130), batteries
Order Notes Each item selected MUST be ordered as a separate line item.
Product Notes All Clamp-on Meters include Test Leads (where applicable) and a carrying case.
AC Current Range 200/600A
AC Current Resolution 0.1A
AC Current Frequency characteristics 40~1kHz
AC Current Method of detection Mean Value
AC Voltage Measurement O
DC Voltage Measurement -
Continuity Check Measurement O
Frequency Measurement -
Temperature Measurement -
Recorder output Function -
Waveform monitor output Function -
Data hold Function O
Auto hold Function -
Peak Hold Function -
Maximum value memory Function -
Mean value display Function -
Filter Function -
Range Hold Function O
Auto ZERO Adjust Function -
Auto power-off Function -