Error(Code:916) Occurs When Installing a 701250 Module in DL850 or SL1000 but Operates Fine in DL750

There are some modules that can be used with the DL750 Scopecorder but cannot be used with the DL850 Scopecorder. This is due to the nature of the FPGA for that module, which cannot be interchanged between the DL750 Scopecorder and the DL850 Scopecorder.

During initialization, the DL850 verifies the serial number, firmware version, and FPGA version of the 701250 Analog Voltage Input Module. If the firmware is not current or if the DL850 Scopecorder finds an incompatible module, it issues an ERROR 916 to alert the user. The SL1000 High Channel Count ScopeCorder experiences a similar issue.

Please contact your local Yokogawa Service Center to arrange the servicing of your module. Analog Voltage Input 701250 and 701255 modules with serial numbers within the range 91F815648 - 91G328674 may be affected.

Related Products & Solutions

Data Acquisition (DAQ)

Simplify complex testing with DAQ systems that flex to your needs. Capture and analyze any physical or electrical signal with ease, transforming signals into actionable insights that speed innovation and improve reliability.

High Speed Data Acquisition

  • PC-based, streaming, local or remote operation
  • 20+ modules, isolated and versatile inputs
  • Up to 200 MS/s or 640 ch

Isolated Oscilloscopes | ScopeCorders

  • An integrated measurement system for every electromechanical application
  • Modular platform combines oscilloscope and DAQ functionality 
  • Capture high-speed transients and low-speed trends 

Oscilloscopes

Accelerate design and troubleshooting with oscilloscopes that combine high-speed sampling, long record lengths, and mixed-signal analysis. From quick debug to advanced R&D, Yokogawa oscilloscopes help you see deeper, work faster, and design smarter.

SL1000 High-Speed Data Acquisition Unit

Predecessor to SL2000. A PC-based ScopeCorder system capable of synchronizing multiple chassis.

Precision Making

Top