How do I Perform a Shunt Calibration With the ScopeCorder Strain Gage Module?

Resources:

To perform a shunt calibration using the 701271 strain gage module and the ScopeCorder, please complete the following procedures:
  1. Connect the strain module 701271, bridge head 701957/701958 and the strain gauge 
  2. Attach the shunt resistor to the bridge head
  3. Set the range unit, range, excitation voltage, and gauge factor
  4. Press the Exec Balance soft key
  5. Enter the reference strain value corresponding to the attached resistor for P2:Y
  6. Select the Unit
  7. Execute the Shunt Cal to perform the calibration
  8. After the calibration is done, the P1:X and P2:X values as well as the Upper and Lower Scale will have changed
  9. If negative strain direction is selected, the P2:Y value must be negative and the P2:X value will also be negative after calibration

Please refer to the attached file for more details on shunt calibration.

Produtos e Soluções Relacionadas

SL1000 High Channel Count ScopeCorder

A headless ScopeCorder system capable of connecting multiple chassis into one PC-based system.

 

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.

Aquisição de Dados (DAQ)

Os sistemas de aquisição de dados da Yokogawa proporcionam a você mais flexibilidade e poder para medir, exibir, armazenar e até mesmo atuar qualquer número de fenômenos físicos ou elétricos.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Osciloscópios

Osciloscópios digitais de classe mundial da Yokogawa: Os osciloscópios digitais têm amostragem de alta velocidade e uma ampla gama de larguras de banda que podem ser utilizadas para projeto e desenvolvimento de dispositivos eletrônicos. As ScopeCorders têm as vantagens de um osciloscópio digital e de um gravador de dados multicanal.

OSCILÓGRAFOS

  • Flexible, high-performance Yokogawa ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

 

Precision Making

Topo