There is no specification of the lowest current for our current probe.
However, it is possible to calculate the lowest current from the accuracy of this probe.
Models 701929/701933/701928/701932
Amplitude Accuracy: +/- ( 1.0% of reading + 1mV)
Output Voltage Rate: 0.1V/A
There is a offset error of +/- 1mV. Even if the input current is zero A, there is a possibility that the output is 1mV.
The scaled current value for a 1mV signal is 10mA. Input current less than 10mA cannot be measured.
If the current being measured is 20mA, the measured deviation will be 20mA +/- 10mA, and the accuracy is 50%.
Therefore, in most cases 20mA cannot be measured.
The lowest current that can be measured accurately is 1A. The measurement accuracy for 1A input current is +/-2% or 1A +/-20mA( 1.0% of reading + 10mA).
Model 701930
Amplitude Accuracy: +/- ( 1.0% of reading + 1mV)
Output Voltage Rate: 0.01V/A
There is a offset error of +/- 1mV. The scaled current value for a 1mV signal is 100mA.
The lowest current that can be measured correctly is 10A.
The measurement accuracy for 10A input current is +/- 2% or 10A +/-200mA( 1.0% of reading + 100mA).
Model 701931
Amplitude Accuracy: +/- ( 1.0% of reading + 5mV)
Output Voltage Rate: 0.01V/A
There is a offset error of +/- 5mV. The scaled current value for a 1mV signal is 500mA.
The lowest current that can be measured correctly is 50A.
The measurement accuracy for 50A input current is +/- 2% or 50A +/-1A( 1.0% of reading + 500mA).
Osciloscópios digitais de classe mundial da Yokogawa: Os osciloscópios digitais têm amostragem de alta velocidade e uma ampla gama de larguras de banda que podem ser utilizadas para projeto e desenvolvimento de dispositivos eletrônicos. As ScopeCorders têm as vantagens de um osciloscópio digital e de um gravador de dados multicanal.