Why is There Multiple Timestamp Rows in the ASCII Header File *.HDR?

Resources:

When viewing the ASCII Header File *.HDR, each timestamp row corresponds to the time a trigger occurs. Multiple timestamp rows will be created if the trigger mode is set to Single(N) and History is set to "All." The DL750 Scopecorder will save the trigger times of N waveforms into separate timestamp rows in the ASCII header file.

If the record length is set short on the DL750 Scopecorder , the memory is divided into blocks, and waveform data is saved into individual blocks each time a trigger occurs. In sequential store function, Single(N) trigger mode, you can specify the waveform acquisition count (block number), and the measurement stops when the specified count is reached.


Please download the sample HDR file, and open it using Notepad for an example of multiple timestamp rows.

 

 

Produtos e Soluções Relacionadas

Data Acquisition (DAQ)

Simplify complex testing with DAQ systems that flex to your needs. Capture and analyze any physical or electrical signal with ease, transforming signals into actionable insights that speed innovation and improve reliability.

High Speed Data Acquisition

  • PC-based, streaming, local or remote operation
  • 20+ modules, isolated and versatile inputs
  • Up to 200 MS/s or 640 ch

Isolated Oscilloscopes | ScopeCorders

  • An integrated measurement system for every electromechanical application
  • Modular platform combines oscilloscope and DAQ functionality 
  • Capture high-speed transients and low-speed trends 

Oscilloscopes

Accelerate design and troubleshooting with oscilloscopes that combine high-speed sampling, long record lengths, and mixed-signal analysis. From quick debug to advanced R&D, Yokogawa oscilloscopes help you see deeper, work faster, and design smarter.

Precision Making

Topo