Why is There Multiple Timestamp Rows in the ASCII Header File *.HDR?

Resources:

When viewing the ASCII Header File *.HDR, each timestamp row corresponds to the time a trigger occurs. Multiple timestamp rows will be created if the trigger mode is set to Single(N) and History is set to "All." The DL750 Scopecorder will save the trigger times of N waveforms into separate timestamp rows in the ASCII header file.

If the record length is set short on the DL750 Scopecorder , the memory is divided into blocks, and waveform data is saved into individual blocks each time a trigger occurs. In sequential store function, Single(N) trigger mode, you can specify the waveform acquisition count (block number), and the measurement stops when the specified count is reached.


Please download the sample HDR file, and open it using Notepad for an example of multiple timestamp rows.

 

 

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

ScopeCorders

  • Flexible, high-performance Yokogawa Test&Measurement ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

Precision Making

Top