701261 Temperature & Universal Voltage Module

701261 Universal Module (2 ch)
Thermocouple (K, E, J, T, L, U, N, R, S, B, W, iron-doped gold/chromel)

How To Check Input Voltage

Standard operating conditions: After a 30-minute warm-up and after calibration; Temp- 23 degrees C +/- 5 degrees C; Humidity- 55% =/-10% RH

Vertical resolution (voltage measurement) during voltage input: 2400 LSB/range; Input connector: Binding post; Effective measurement range: Twice of setting range

Voltage and temperature formula: Temperature coefficient- [Voltage Measurement] Zero point +/-(0.01% of range)/degrees C (typical); Gain +/-(0.02% of range)/degrees C (typical)

701261 / 701262 Universal (Voltage/Temp.) Module – specifications
Item Specifications
Standard operating conditions Temperature: 23 ±5 °C
Humidity: 20 to 80 % RH
After a 30 minute warm-up and after calibration
Function Temperature (thermocouple) or voltage measurement (switchable)
Effective measurement range [Voltage measurement] 20 div (display range: 10 div)
Number of input channels 2
Input coupling settings TC, DC, AC, GND
TC: Temperature (thermocouple) measurement
DC: Voltage measurement (DC coupling)
AC: Voltage measurement (AC coupling)
When measuring voltage – maximum sampling rate 100 kS/s
Temperature measurement data update rate 500 Hz
Input type Isolated unbalanced
Measurement range/accuracy1 [Voltage measurement] Voltage scale: 5 mV/div to 20 V/div (1-2-5 steps); Voltage accuracy: ±(0.25 % of 10 div)
[Temperature measurement]2
Thermocouple standards: K, E, J, T, N, R, S, B: IEC 60584-1, DIN IEC 60584-1, JIS C1602; W: W-5 %Re/W-26 %Re (Hoskins Mfg. Co.), ASTM E988; L: Fe-CuNi, DIN43710; U: Cu-CuNi, DIN43710
Type Measurement range Accuracy
K −200 to 1300 °C ±(0.1 % of reading + 1.5 °C)
E −200 to 800 °C Except ±(0.2 % of reading + 1.5 °C) for −200 °C to 0 °C
J −200 to 1100 °C  
T −200 to 400 °C  
L −200 to 900 °C  
U −200 to 400 °C  
N 0 to 1300 °C  
R 0 to 1700 °C ±(0.1 % of reading + 3 °C)
S 0 to 1700 °C Except, 0 to 200 °C: ±8 °C
200 to 800 °C: ±5 °C
B 400 to 1800 °C ±(0.1 % of reading + 2 °C)
Except, 400 to 700 °C: ±8 °C
Effective range is 400 to 1800 °C
W 0 to 2300 °C ±(0.1 % of reading + 3 °C)
Au7Fe3 0 to 300 K 0 to 50 K: ±4 K
50 to 300 K: ±2.5 K
Frequency characteristics1 (−3 dB point when a sine wave of amplitude ±3 div is applied) [Voltage measurement] DC to 40 kHz
[Temperature measurement] DC to 100 Hz
Maximum input voltage4 (at a frequency of 1 kHz or less) Both temperature and voltage input: 42 V (DC+ACpeak) (as a value that meets the safety standard)
150 V (DC+ACpeak) (maximum allowable voltage, as a value that does not damage the instrument when applied)
Maximum rated voltage to ground5 – between input and case and between inputs (at a frequency of 1 kHz or less) Both temperature and voltage input: 42 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical resolution [Voltage measurement] During voltage input: 2400 LSB/div
[Temperature measurement] When measuring temperature: 0.1 °C
−3 dB point when AC coupled [Voltage measurement] 0.5 Hz or less
Input connector Binding post
Input impedance Approx. 1 MΩ
Common mode rejection ratio [Voltage measurement] 80 dB or more (50/60 Hz) (typical value6)
[Temperature measurement] 120 dB or more (50/60 Hz, with 2-Hz filter ON, signal source resistance of 500 Ω or less) (typical value6)
Residual noise level (input section shorted) [Voltage measurement] ±100 µV or ±0.01 div whichever is greater (typical value6)
A/D converter resolution [Voltage measurement] 16 bits (2400 LSB/div)
Temperature coefficient [Voltage measurement] Zero point: ±(0.01 % of 10 div)/°C (typical value6)
Gain: ±0.02%/°C (typical value6)
Reference junction compensation accuracy (when the input terminal temperature is balanced) K, E, J, T, L, U, N: ±1 °C
R, S, B, W: ±1.5 °C
Au7Fe: ±1 K
Bandwidth limit [Temperature measurement] (Digital filter + analog filter): Select from Full, 30 Hz, 8 Hz, and 2 Hz + 150 Hz secondary analog filter
[Voltage measurement] Select from Full, AUTO, 4 kHz, 400 Hz, 40 Hz
Cutoff characteristics: −12 dB/OCT (typical value,6 setting other than AUTO)
Cutoff frequency (fc) when set to AUTO (701262 only)
Sample rate Cutoff frequency (fc)
100 kS/s or more 40 kHz
100 S/s to 50 kS/s 40 % of the sample rate
50 S/s or less 20 Hz
Cutoff characteristics for AUTO: −65 dB at 2 × fc (typical value6)
Table of cutoff frequency characteristics of the anti-aliasing filter (AAF) When the filter is set to Auto, the anti-aliasing filter and low-pass filter are automatically set according to the sample rate.
Sample rate AAF Low-pass filter
100 kS/s 40 kHz OFF
50 kS/s 20 kHz OFF
20 kS/s 8 kHz OFF
10 kS/s 4 kHz 4 kHz
5 kS/s 2 kHz 4 kHz
2 kS/s 800 Hz 4 kHz
1 kS/s 400 Hz 400 Hz
500 S/s 200 Hz 400 Hz
200 S/s 80 Hz 400 Hz
100 S/s 40 Hz 40 Hz
50 S/s 20 Hz 40 Hz
20 S/s to 5 S/s 20 Hz 40 Hz
2 S/s or less 20 Hz 40 Hz
Ext sample 40 kHz OFF
Wiring / measurement points
Voltage terminal diagram (H, L) with measurement points 4 and 5

1. Value measured under standard operating conditions.
2. Does not include the reference junction temperature compensation accuracy.
3. This module supports Au7Fe with 0.07 % metal content with respect to gold.
6. Typical values represent typical or average values. They are not strictly warranted.

DL950 ScopeCorder

Capture and analyze a wide variety of electromechanical signals and serial buses. High sample rate, long recording times, advanced triggers, and real-time analysis.

DL350 Portable ScopeCorder

A compact, battery-powered, touch screen instrument you can take into the field featuring the measurement and recording capabilities you need for maintenance and troubleshooting.

DL850E/DL850EV ScopeCorder

An all-in-one electromechanical measurement system offering the deepest measurement and analysis features to turn data into insights.

DL850/DL850V ScopeCorder

Combining the capabilities of an Oscilloscope and the characteristics of a Data Acquisition Recorder, the Yokogawa Scopecorder is a versatile tool for engineers making measurements in the industrial, transportation and energy sectors.

SL1000 High-Speed Data Acquisition Unit

Predecessor to SL2000. A PC-based ScopeCorder system capable of synchronizing multiple chassis.


 

Looking for more information on our people, technology and solutions?


Contact Us

Precision Making

Top