技术与服务热线:400 820 0372

Why is There Multiple Timestamp Rows in the ASCII Header File *.HDR?

Resources:

When viewing the ASCII Header File *.HDR, each timestamp row corresponds to the time a trigger occurs. Multiple timestamp rows will be created if the trigger mode is set to Single(N) and History is set to "All." The DL750 Scopecorder will save the trigger times of N waveforms into separate timestamp rows in the ASCII header file.

If the record length is set short on the DL750 Scopecorder , the memory is divided into blocks, and waveform data is saved into individual blocks each time a trigger occurs. In sequential store function, Single(N) trigger mode, you can specify the waveform acquisition count (block number), and the measurement stops when the specified count is reached.


Please download the sample HDR file, and open it using Notepad for an example of multiple timestamp rows.

 

 

相关产品和解决方案

数据采集设备

Yokogawa数据采集系统具备出众的灵活性和性能,能测量、显示、存储甚至促动各种物理和电气现象。

波形测量仪器

Yokogawa在业界首推示波记录仪的概念,将数字示波器和多通道数据记录仪的优点集于一身。横河示波记录仪作为世界一流的示波记录仪具有高速采样能力和大测量带宽的优点,可广泛应用于电子工程的设计和开发过程。

横河示波器既有面向低成本市场的4CH MSO示波器DLM3000系列,也有业界唯一的8CH MSO示波器DLM4000系列,在由智能控制的消费电子和工业驱动等电路设计的众多领域中,成为全球工程师的首选工具。

示波记录仪

我们的示波记录仪是灵活、强大的多通道测试和测量解决方案,将高精度示波器和传统有纸图表记录仪的优点集于一身。Yokogawa的示波记录仪除了能在它的大显示屏上显示信号之外,还可以将信号保存到记录纸和/或存储介质中以便长期储存(如30天或更长),此外它还可以捕捉并分析非常快的瞬变信号。

高速数据采集器和数采仪

Yokogawa高速数据采集系统提供业界领先的隔离、位分辨率、采样速率和存储长度,带有独立的通道硬件和使用简单的软件。是示波记录仪中最领先的技术。

Precision Making

返回顶部
WeChat QR Code
微信扫一扫
获得更多专业服务