Why Won't the Measured Values of the Fundamental Wave Components and the Bar Graph of the Harmonics Display on the WT3000?

When the selected data update cycle of the fundamental wave is shorter than the width of the analysis window determined by the fundamental frequency (cycle of the fundamental wave), measurement is not performed and no data is displayed. Change the currently selected data update rate to something longer.

For harmonic measurement of a 50 Hz distorted wave signal, the fundamental frequency is 50 Hz and the width of the analysis window is ten waves, so the data measurement interval is 200 ms. Since (data measurement interval + data computation interval) = approximately 300 ms or more, select a data update rate of 500 ms or more.

 

Produtos e Soluções Relacionadas

Precision Power Analyzer WT3000

With 0.02% accuracy and 1MHz bandwidth, the WT3000 delivers where the highest precision measurements are required. It is the industry standard for R&D work on inverters, motor drives, lighting systems and electronic ballasts, UPS systems, aircraft power, transformer testing, and other power conversion devices.