OSA: Spatial Light Measurement

Download (197.0 KB)

When measuring light emitted into space, the emitted light needs to be captured into an optical fiber before entered in to an optical spectrum analyzer (OSA). However, there are cases where ordinary optical fibers cannot capture a sufficient amount of light in the OSA, such as the output of a light source that emits light from a surface such as an LED or a fluorescent lamp and pulsed light with a low duty ratio (short emission time and long repetition period).

The applicable models have a free space structure inside the optical input port which allows to use an optical fiber with a core diameter of up to 800 µm, so the use of large core fiber helps the OSA collect more light. Please note that the actual minimum wavelength resolution is limited by the core diameter of the optical fiber used (See lower right table) and for wavelengths over 2000 nm, an optical fiber with low infrared absorption (e.g. fluoride fiber) is required. In OSA that uses an optical fiber in the optical input port, even if the core diameter of the external optical fiber is increased, the amount of light that can be taken in is not increased because it is limited by the core diameter of the internal optical fiber.

 

Core diameter dependence (Reference 50 μm)

Core diameter dependence (Reference 50 μm)

 

How optical input structure affects optical power measurement

How optical input structure affects optical power measurement

 

Limitation of minimum wavelength resolution by optical fiber

Optical fiber

Minimum wavelength resolution (nm)

Type

Core

(μm)

AQ6360

Telecom

AQ6370

Telecom

AQ6373

VIS

AQ6374

VIS- exNIR

AQ6375

exNIR-SWIR

AQ6376

exNIR-MWIR

AQ6377

SWIR-MWIR

SM

5

0.1

0.02

0.02 *1

0.05

0.05

0.1

SM

5

0.2

SM

9.5

0.1 *1

0.02 *1

0.02

0.05* 1

0.05 *1

0.1 *1

SM

9.5

0.2

GI

50

0.1

0.05

0.05

0.2

0.1

0.1

SM

13

0.2 *1

GI

62.5

0.1

0.05

0.05

0.2

0.1

0.1

MM

50

0.5

SI

80

0.2

0.1

0.1

0.2

0.2

0.2

MM

80

1

SI

100

0.5

0.2

0.2

0.5

0.5

0.5

MM

100

1

SI

200

1

0.5

0.5 *2

0.5 *2

1

1

MM

200

2

SI

400

2

1

1 *2

1 *2

2

2

MM

400

5

SI

800

2

2

2 *2

2 *2

2

2

 

 

 

Approximate values at reference wavelengths   *1 Reference optical fiber type *2 Core size setting: Large

Related Industries

Related Products & Solutions

AQ6361 Telecom Production 1200 - 1700 nm

  • 0.03 nm resolution
  • ±20 pm accuracy
  • 65 dB close-in dynamic range
  • -80 dBm level sensitivity
  • Highest sweep speed
  •  

AQ6370E Telecom 600 - 1700 nm

  • 0.02 nm resolution
  • ±8 pm accuracy
  • 78 dB close-in dynamic range
  • -90 dBm level sensitivity
  • Single-mode and multi-mode
  •  

AQ6373E Visible Wavelength 350 - 1200 nm

  • 0.02 nm resolution
  • ±50 pm accuracy
  • 60 dB close-in dynamic range
  • -80 dBm level sensitivity
  • Single-mode and multi-mode
  • Industrial & biomedical laser testing

AQ6374E Wide Wavelength Range 350 - 1750 nm

  • 0.05 nm resolution
  • ±50 pm accuracy
  • 60 dB close-in dynamic range
  • -80 dBm level sensitivity
  • Single-mode and multi-mode
  • Visible to communications wavelengths

AQ6375E Long Wavelength 1000 - 2500 nm

  • 0.05 nm resolution
  • ±50 pm accuracy
  • 55 dB close-in dynamic range
  • 90 dB measurement dynamic range
  • -70 dBm level sensitivity
  • Single-mode and multi-mode
  • Advanced pulsed light measurement (APLM) mode

Optical Spectrum Analyzers

  • High-resolution measurement of optical power vs wavelength
  • Ideal for R&D, manufacturing, and laser characterization
  •  

Precision Making