Yokogawa CA450 includes specialised functions for troubleshooting, maintenance and repair of plant equipment
With the loop check function, designed for tests on 2-wire transmitters, the CA450 supplies 24 VDC power to the transmitter and measures the DC output in milliamperes to an accuracy of 0.05% up to 30 mA. A HART mode with loop power and 250 ohm resistance allows HART and BRAIN communications to be tested. A transmitter simulator function allows transmitter I/O to be checked by drawing in ('sinking') designated currents from the external voltage generators of the instrumentation. The 4-20 mA generator output includes span, step, auto-step and sweep capabilities.
The CA450 can also be used to carry out operating checks and adjustments on valve positioners, including zero span adjustment, response testing using the auto-step function, and tests for hunting, sticking/slipping and valve open/close errors. Controller programs can also be verified using the CA450's 4-20 mA step generation function, which switches between 4 and 20 mA in 4mA output steps.
Electrical measurements include power supply peak-to-peak measurement with instantaneous detection in the DC voltage measurement mode, and AC/DC current measurement up to 180 A AC and 3000 A AC in sensor mode.
The CA450 can also be used to check panel meters for correct connection, reproducibility and errors.
The CA450 features a safe design with current terminal shutters to protect both the operator and the instrument against incorrect connection. It conforms to the 600 V CAT.4 and 1000 V CAT.3 standards. An optional adaptor allows the instrument to be linked with a PC, with data transferred via an infrared USB interface.