Is There a Way to Recover a DL850 WDF File That Cannot be Opened in Xviewer?

If the DL850 WDF file was saved through the hard disk recording feature (HD Recording), then the file may be recovered by using the Finalize feature in Xviewer File Utility. Files that were saved through the normal save procedure or dual capture feature cannot be recovered.
 
To recover the corrupted file, please complete the following procedures:
  1. Open Xviewer
  2. Click on Tools ⇒ DL850 Advanced Utility ⇒ File Utility....
  3. Click on the Finalize tab
  4. Click and Browse and Open the corrupted file
    (File should display in the file list)
  5. Verify the status of the corrupted file underneath the Finalize column. Files denoted by "Undone" can be recovered. Files denoted by "Done" cannot be recovered.
  6. Select whether to Overwrite the original file
  7. Click on Execute

For more information on file Finalize and Xviewer File Utility, please refer to the JS01 Xviewer Advanced File Utility Tutorial.

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

World-class digital oscilloscopes from Yokogawa: The digital oscilloscopes have high-speed sampling and a wide range of bandwidths that can be utilized for design and development of electronic devices. The ScopeCorders have the advantages of both a digital oscilloscope and a multi-channel data recorder.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.

Top