OSA: Characterization of Optical Amplifier Gain and Noise Figure

Download (398.5 KB)

Applicable model*: AQ6380/AQ6370 series/AQ6360

With the increase in communication traffic, the importance of optical amplifiers such as EDFAs (erbium-doped fiber amplifiers) that can directly amplify optical signals without converting them into electrical signals is increasing in backbone optical transmission networks.

The applicable models are equipped as standard with an optical amplifier analysis function (EDFA-NF) that automatically calculates the gain and noise figure (NF) of the optical amplifier from the input and output optical spectrum of the optical amplifier.

Since the Amplified Spontaneous Emission (ASE) generated by the optical amplifier is superimposed on the output light of the optical amplifier, it is important to measure this noise component separately in optical amplifier evaluation. The optical amplifier analysis function identifies the ASE component by the curve fitting and interpolation method. In addition, the curve fitting method and analysis conditions can be set according to the actual spectrum, so the gain and NF can be obtained accurately.

* All versions unless otherwise specified. 

  AQ6373 family are not applicable.

Optical amplifier analysis function (EDFA-NF)

nCalculate the gain and NF for each channel of the input optical signal.
(Compatible with DWDM signals.)
nAnalysis results such as wavelength, gain, and NF are displayed in a list or graph for each channel.
nPossible to set analysis parameters such as channel detection and ASE level detection

OSA: Characterization of Optical Amplifier Gain and Noise Figure

Related Products & Solutions

AQ6361 Telecom Production 1200 - 1700 nm

  • 0.03 nm resolution
  • ±20 pm accuracy
  • 65 dB close-in dynamic range
  • -80 dBm level sensitivity
  • Highest sweep speed
  •  

AQ6370E Telecom 600 - 1700 nm

  • 0.02 nm resolution
  • ±8 pm accuracy
  • 78 dB close-in dynamic range
  • -90 dBm level sensitivity
  • Single-mode and multi-mode
  •  

AQ6373E Visible Wavelength 350 - 1200 nm

  • 0.02 nm resolution
  • ±50 pm accuracy
  • 60 dB close-in dynamic range
  • -80 dBm level sensitivity
  • Single-mode and multi-mode
  • Industrial & biomedical laser testing

AQ6374E Wide Wavelength Range 350 - 1750 nm

  • 0.05 nm resolution
  • ±50 pm accuracy
  • 60 dB close-in dynamic range
  • -80 dBm level sensitivity
  • Single-mode and multi-mode
  • Visible to communications wavelengths

AQ6375E Long Wavelength 1000 - 2500 nm

  • 0.05 nm resolution
  • ±50 pm accuracy
  • 55 dB close-in dynamic range
  • 90 dB measurement dynamic range
  • -70 dBm level sensitivity
  • Single-mode and multi-mode
  • Advanced pulsed light measurement (APLM) mode

Optical Component Test System

  • Modular platform for automated optical testing
  • Attenuators, switches, power meters, light sources, and SMUs