Is There any Dead Time Between Files When Recording Multiple Files From the SL1000?

Depending on the mode, the dead time may range from 0 to 10ms approximately. The dead time for the different modes are listed below.
  • Free Run Mode: 0ms
  • Normal Mode without HDD recording: 10ms
  • Normal Mode with HDD recording: 10ms + HDD recording time
  • Single(N) Mode: 10ms 

Another important concern is whether there is any data loss between multiple files, in which case the answer is no. The SL1000 allocates enough internal memory to buffer incoming data for save operations and therefore, data is not lost.

Related Products & Solutions

SL1000 High Channel Count ScopeCorder

A headless ScopeCorder system capable of connecting multiple chassis into one PC-based system.

 

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

World-class digital oscilloscopes from Yokogawa: The digital oscilloscopes have high-speed sampling and a wide range of bandwidths that can be utilized for design and development of electronic devices. The ScopeCorders have the advantages of both a digital oscilloscope and a multi-channel data recorder.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.

 
Top