Can the GS610 or GS820 be Used as a DMM to Only Measure Voltage?

No, the GS610 and the GS820 cannot be used solely as a DMM to measure only voltage. This is because the Sense terminal cannot be separated from the Output terminal on the instrument. The GS Series uses the Output terminal to source and measure the DUT, it is NOT using the Sense terminal for measurement. The Sense terminal is used only in the case of the 4-wire mode.  
  • In 2-wire mode, the Sense terminal is internally connected to the Output terminal and cannot be used.
  • In 4-wire mode, the Sense terminal is internally connected to the Output terminal by a 40kΩ resistor.
 
For the single channel GS610, the Output and Sense terminal must be connected at the same point. For the multi-channel GS820, you can use CH1 and CH2 to measure separate parts of the DUT.


Please refer to the attached PDF for a diagram of the 2-wire and 4-wire connection method.

Related Products & Solutions

GS200 DC Voltage / Current Source

  • Ultra low noise (3 µVp-p)
  • Ranges: ±30 V, ±200 mA
  • Resolution: 100 nV / 10 nA

GS610 Source Measure Unit

  • 1-Channel
  • ±110V / ±0.5A or ±12V / ±3.2A
  • Resolution: 1 µV / 100 pA
  • Sweep: Linear, Log, Custom
  • Output: DC, Pulse (50 µs to 3600 s)

GS820 Source Measure Unit

  • 2-Channel
  • ±50 V / ±1.2 A or ±18V / ± 3.2 A
  • Resolution: 1 µV / 1 pA
  • Sweep: Linear, Log, Custom
  • Output: DC, Pulse (50 µs to 3600 s)

Signal Generators, Sources and Supplies

Sources and signal generators offer high accuracy and functionality for standalone use or as key components in fast test systems.

Source Measure Units

A Source Measurement Unit (SMU) is a flexible test instrument which is capable of both sourcing and measuring voltage and/or current at the same time.

Precision Making

Top