Why Can't the Digital Oscilloscopes DL9000 / SB5000 Display Harmonics?

Please make sure to set Time/Div to a setting which will display a minimum of sixteen (16) Cycles.     
 
To setup the instrument for harmonic analysis:
  1. Press the WINDOW1 key
  2. Press the Type soft key
  3. Select Harmonics
  4. Turn the T/Div knob until it is at least 16 cycles (IEC harmonics requires 16 cycles minimum)    
               
        

Related Products & Solutions

DL9000 DSO Series

500MHz, 1.0GHz, and 1.5GHz DSOs for debug and high performance applications. 10th
generation oscilloscope from Yokogawa with industry leading 2.5 million wfms/sec and lowest dead time. Winner of Test & Measurement World's "Best in Test" award.

Mixed Signal Oscilloscopes

  • Yokogawa mixed signal oscilloscopes
  • Simultaneous, time correlated observations
  • Analysis of analog with digital (logic) signals
  • Troubleshooting electrical anomalies, measuring parametric values, monitoring signals

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

Precision Making

Top