What is the Sampling Interval of the WE7241 Digital Thermometer Module?

The sampling interval changes with the type of time base and the number of measurement channels per module. There are two possible methods: 
  1. If Time base = internal
    Number of channels : 1-2 = 0.2 s, 3-5 = 0.3 s, 6-7 = 0.4 s, 8-10 = 0.5 s. (When modules are connected in a series, the sampling interval is that of the slowest module in the series.)  
  2. If Time base = external
    2 s or more (regardless of the number of channels).

Related Products & Solutions

Data Acquisition (DAQ)

Simplify complex testing with DAQ systems that flex to your needs. Capture and analyze any physical or electrical signal with ease, transforming signals into actionable insights that speed innovation and improve reliability.

High Speed Data Acquisition

  • PC-based, streaming, local or remote operation
  • 20+ modules, isolated and versatile inputs
  • Up to 200 MS/s or 640 ch

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.