Why Can't the Width level of the Window Trigger on the Digital Oscilloscope DL1500 Series be set to Below the Voltage Sensitivity?

The width level of the Window trigger cannot be set to a value less than the voltage corresponding to 1 division at the specified voltage sensitivity. For example, if the voltage sensitivity is set to 500 V/div, then the width level cannot be set below 500 V. This limitation is due to the fact that the trigger sensitivity of the Digital Oscilloscope DL1500 Series is 1 Divp-p.

Related Products & Solutions

Mixed Signal Oscilloscopes

  • Analyze analog and digital signals simultaneously
  • Advanced triggering and high-speed waveform capture
  • Power analysis, serial bus analysis, & switching loss
  •  
  •  
  •  

Oscilloscopes

Accelerate debugging and gain deeper insight with high-resolution oscilloscopes designed for speed, clarity, and precision.

Precision Making

Top