Are the Standard Operating Conditions for the DL850 Equivalent to the Compensated Temperature Range?
No, the Standard Operating Conditions are not equivalent to the compensated temperature range. The unit's accuracy specifications are valid when operated within the Standard Operating Conditions and when the temperature of the unit does not change. The temperature coefficient needs to be applied when the temperature of the unit changes even if the temperature is in the Standard Operating Conditions.
Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.