High-Voltage 100 kS/s sample rate, 16-Bit resolution, 40 kHz bandwidth, 2 channels, Isolated, 850 V*3 Maximum input voltage, ±0.25% DC accuracy, with RMS, and high noise immunity.
ITEM |
SPECIFICATIONS |
Input channels |
2 |
Input type |
Isolated unbalanced |
Input coupling |
AC, DC, GND, AC-RMS, and DC-RMS |
Input connector |
Plug-in terminal (safety terminal) |
Input impedance |
1 MΩ±1%, approx. 35 pF |
Maximum sample rate |
100 kS/s |
Frequency range |
(-3dB)*1 |
Waveform observation mode |
DC to 40 kHz |
RMS observation mode |
DC, 40 Hz to 10 kHz |
A/D conversion resolution |
16-bit (2400 LSB/div) |
Voltage-axis sensitivity setting |
20 mV/div to 200 V/div (1-2-5 steps) |
Maximum input voltage (1 kHz or less) |
|
In combination with 758933+701954 (1:1) |
850 V (DC+ACpeak) |
Direct input |
850 V (DC+ACpeak) |
Maximum allowable common mode voltage (1 kHz or less) |
|
In combination with 758933+701954 (1:1)*9 |
H side: 700 Vrms (CAT @), L side: 400 Vrms (CAT @) |
Direct input H/L sides: |
30 Vrms (42 VDC+ACpeak)*11 |
Vertical (voltage) axis accuracy*1 |
|
Waveform observation mode |
DC accuracy ±(0.25% of 10 div) |
RMS observation mode |
DC accuracy ±(1.0% of 10 div) |
AC accuracy (sinewave input) |
±(1.5% of 10 div) At frequency of 40 Hz to 1 kHz |
AC accuracy (crest factor 2 or less) |
±(2.0% of 10 div) At frequency of 40 Hz to 1 kHz |
AC accuracy (crest factor 3 or less) |
±(3.0% of 10 div) At frequency of 40 Hz to 1 kHz |
Temperature coefficient (Waveform observation mode) |
|
Zero point |
±(0.02% of 10 div)/: (Typ.) |
Gain |
±(0.02% of 10 div)/: (Typ.) |
Bandwidth limit |
OFF/100 Hz/1 kHz/10 kHz |
Response time (RMS observation mode) |
|
Rising |
(0 to 90% of 10 div) 100 ms (Typ.) |
Falling |
(100 to 10% of 10 div) 250 ms (Typ.) |
Weight |
Approx. 250g |