AQ6380 Highest Performance Optical Spectrum Analyzer 1200 - 1650 nm and 900 - 1650 nm

AQ6380 Optical Spectrum Analyzer | 2022 Lightwave Innovation Reviews | Yokogawa Test&Measurement

The AQ6380 OSA is the highest performance optical spectrum analyzer from Yokogawa Test&Measurement. Its excellent optical wavelength resolution, accuracy, and close-in dynamic range specifications allow optical signals in close proximity to be clearly separated and precisely measured. 

This OSA incorporates smart technology and functionality including an intuitive touchscreen, automated wavelength calibration, and optimized sweep speed, allowing users to operate more efficiently. In addition, the AQ6380 OSA includes gas purging mechanisms, fully-automated wavelength calibration, compatibility with high-resolution and high sample counts, and single-mode fiber input.

Key Features:

  • High wavelength resolution: 5 pm (0.624 GHz)
  • High wavelength accuracy: ±5 pm 
  • Wide close-in dynamic range: 65 dB 
  • High stray light suppression: 80 dB 
  • Wavelength coverage: 1200 - 1650 nm (standard model), 900 - 1650 nm (extended model)

 

Are you involved with Optical Testing for learning?
Click here to learn about our special OSAs for Education.

Free Yokogawa Fiber-Optic Communication Poster

AQ6380 OSA Optical Spectrum Analyzer with 5 pm Resolution | Yokogawa Test&Measurement

Introducing the new flagship AQ6380 Optical Spectrum Analyzer from Yokogawa Test&Measurement that boasts an unprecedented 5 pm resolution! Features of the AQ6380 monochromator OSA include: -- 5 pm high wavelength resolution -- ±5 pm wavelength accuracy -- 1200 nm - 1650 nm wavelength range -- 65 dB wide close-in dynamic range -- 80 dB stray light suppression -- Automated wavelength calibration -- Gas purging to minimize water vapor absorption -- DUT-specific interface and test apps -- Backward-compatible remote interface -- 10.4in capacitive touchscreen with intuitive interface -- Up to 20x faster measurement -- Able to be remotely controlled and emulated over your network Learn more about the Highest Resolution Optical Spectrum Analyzer AQ6380 from Yokogawa Test&Measurement: https://tmi.yokogawa.com/solutions/products/optical-measuring-instruments/optical-spectrum-analyzer/aq6380-optical-spectrum-analyzer/ __________________________________________________________________ Connect with Yokogawa Test&Measurement on LinkedIn: https://www.linkedin.com/showcase/yokogawa-test&measurement Meet the Precision Makers at https://tmi.yokogawa.com/ Yokogawa has been developing measurement solutions for 100 years, consistently finding new ways to give R&D teams the tools they need to gain the best insights from their measurement strategies. The company has pioneered accurate power measurement throughout its history and is the market leader in digital power analyzers. Yokogawa instruments are renowned for maintaining high levels of precision and for continuing to deliver value for far longer than the typical shelf life of such equipment. Yokogawa believes that precise and effective measurement lies at the heart of successful innovation - and has focused its own R&D on providing the tools that researchers and engineers need to address challenges great and small.

Unparalleled Optical Performance


5 pm (0.624 GHz) high wavelength resolution

The AQ6380 achieves a wavelength resolution of 5 pm. It enables to separate closely allocated DWDM channels and modulation side peaks of optical transceivers.

AQ6380 Optical Spectrum Analyzer High Wavelength Resolution | Yokogawa Test&Measurement
Modulated spectrum of 10G optical transceiver
 

65 dB wide close-in dynamic range

The monochromator has sharper spectral characteristics than ever. Signals in close proximity (e.g., residual longitudinal modes of external cavity laser) can be clearly separated and accurately measured.

AQ6380 Optical Spectrum Analyzer Close In Dynamic Range | Yokogawa Test&Measurement
Spectrum of external cavity laser
 

Wide wavelength range and variable resolution support multiple applications

The AQ6380 has a wavelength band of 1200 to 1650 nm. This means a single unit can meet diversifying wavelength measurement needs. Wavelength resolution can be varied from 5 pm to 2 nm, supporting a wide range of applications from narrowband peak/notch measurements to wideband spectral measurements. In addition, the increased resolution contributes to an improvement in measurement speed and accuracy for low power signals.

AQ6380 Optical Spectrum Analyzer Filter | Yokogawa Test&Measurement
Transmission spectrum of optical filter
 

80 dB stray light suppression

Stray light is optical noise caused by the diffuse reflection of incident light inside a monochromator. In situations such as laser SMSR measurement, where multiple optical spectra with different levels are measured at the same time, the stray light can interfere with the measurement. When this occurs, high stray light suppression performance is required. The AQ6380 provides high dynamic range measurements with excellent stray light suppression performance of 80 dB, with no spurious noise generated.

AQ6380 Optical Spectrum Analyzer Stray Light Suppression | Yokogawa Test&Measurement
Stray light suppression performance
 

Gas purging mechanism to minimize water vapor absorption

In the near-infrared wavelength range there are wavelength regions where strong light absorption is observed due to the influence of water vapor in the air. In an OSA, there is a wavelength band in which the light absorption characteristics of water vapor inside the monochromator are noticeably detected. Such phenomena interferes with accurate optical spectrum measurements in the applicable wavelength band. The AQ6380 is equipped with a purge mechanism that replaces the air inside the monochromator with nitrogen or dry air by continuously supplying it through dedicated ports on the back panel. This results in accurate measurements that are unaffected by the light absorption phenomenon of water vapor.

AQ6380 Optical Spectrum Analyzer Gas Purge | Yokogawa Test&Measurement
Effect of purging
 

±5 pm wavelength accuracy

The AQ6380 offers ±5 pm in the C band to meet the most stringent accuracy requirements. It also offers ±10 pm in the S and L bands and ±50 pm over the entire wavelength range. With such accuracy, some applications may not require an optical wavelength meter anymore. Periodic self-wavelength calibration using the built-in wavelength reference light source ensures long-term stability for each measurement.

Automated wavelength calibration maintains high accuracy

Ambient temperature change, vibrations, and shock affect the measurement accuracy of high precision products such as optical spectrum analyzers. The AQ6380 delivers high-precision measurements long-term with the wavelength calibration and alignment adjustment functions using the built-in light source. Wavelength calibration with the internal light source can be performed fully automatically and regularly without an external fiber cord. Wavelength calibration using an external light source is supported and is accomplished by setting the exact wavelength.

Single-mode fiber input

The optical input uses a PC-type single-mode fiber. It achieves an optical return loss of 30 dB or more (even when connected with a PC-type optical plug) and reduces the impact of reflectioni-sensitive optical devices on measurements.

Advancing Market Capabilities
The extreme test requirements of our research called for an OSA with extended MIR spectrum bandwidth capabilities up to 5pm, but we couldn’t find one on the market capable of measuring optical inputs at these wavelengths. Yokogawa Test&Measurement rose to the challenge and developed a new OSA model for us that would. Not only do we now have an instrument that is practically custom-made for our needs, it provides repeatable, accurate, and trusted measurement outputs and is easy to learn and use. Their equipment and ability to create a new optical measurement solution has definitely increased the overall efficiency and productivity of our research team.
— Martin Bernier, PhD, P.Eng.
Full Professor, Centre de Optique, Photonique, et Laser, Université Laval

Improve measurement efficiency and productivity

Up to 20x faster measurement

With a new sensitivity mode (RAPID) specialized for CW light, measurement speed is up to 20x faster than conventional models.

New sensitivity settings optimize measurement time

Sensitivity settings have a significant impact on measurement time. The AQ6380 has two modes and 19 sensitivity settings. Shorten measurement time by selecting the optimum sensitivity according to the type of optical signal and the minimum sensitivity requirement. Measurement sensitivity can also be set numerically. The appropriate sensitivity setting is automatically selected when entering the required minimum sensitivity value.

  • RAPID: Sensitivity mode for fast measurement specialized for CW light
  • TRAD: Traditional sensitivity mode that supports both CW light and pulsed light
RAPID Mode TRAD Mode AQ6380 Optical Spectrum Analyzer | Yokogawa Test&Measurement
Note: The sensitivity is a reference value at 1550 nm, not a guaranteed value

 

Built-in analysis functions eliminate post-processing tasks

The AQ6380 has built-in analysis functions to characterize optical spectrum from a variety of optical systems and devices including WDM, DFB-LD, EDFA, and filters. The automatic calculation of the major parameters of the device under test will contributes to fast characterization. Analysis functions include DFB-LD, FP-LD, LED, spectral width (peak/notch), SMSR, optical power, WDM (OSNR), EDFA (NF and gain), filter (peak/bottom), and WDM filter (peak/bottom).

AQ6380 Optical Spectrum Analyzer Laser Analysis 1nm | Yokogawa Test&Measurement
Example of SMSR analysis

 

DUT-oriented test apps simplify the test process

The appplication mode (APP) transforms an OSA into a versatile machine dedicated to a device under test (DUT). APP mode provides a DUT-specific user interface that navigates the user from configuration settings to test result output without worrying with other OSA settings. The AQ6380 comes pre-installed with several standard applications including WDM testing, DFB-LD testing, and FP-LD testing. Additional optional applications are available for download from the Yokogawa Test&Measurement website.

AQ6380 Optical Spectrum Analyzer App Menu | Yokogawa Test&Measurement
APP menu window
AQ6380 Optical Spectrum Analyzer APP WDM Test | Yokogawa Test&Measurement
WDM test application

 

Backward-compatible remote interface for easy upgrade of test system

Easily build an automated measurement system using a remote-control interface (Ethernet or GP-IB). The remote command set conforms to the Standard Commands for Programmable Instruments (SCPI), compatible with AQ6370 series and AQ6319, as well as proprietary AQ6317-compatible commands, allowing for easy upgrades to existing measurement systems.

OSA Viewer enables emulation and remote control on a PC

Emulate and remotely control the AQ6380 using OSA Viewer, the application software included with the AQ6370 Viewer. OSA Viewer's user interface and analysis capabilities allow R&D and production users to easily view and analyze waveforms on their PC or laptop. The AQ6380 can be controlled remotely via a direct connection or over a network. The screen image from the AQ6380 displays in real time on remotely-connected PCs or laptops, where you can operate it as if you are using the AQ6380 directly, and data files saved on the AQ6380 are remotely transferable to your computer. These features are ideal when R&D users need to evaluate and analyze measurement data and optimize test conditions and troubleshoot on remote lines, as well as when Production users need to collect and analyze measurement results of remote production lines.

Interface

AQ6380 Optical Spectrum Analyzer Interface | Yokogawa Test&Measurement

AQ6380 OSA Optical Spectrum Analyzer Interface | Yokogawa Test&Measurement

Measurement applications and user-friendly multi-touch operation and connectivity

With over four decades perfecting the user experience, the Precision Makers of Yokogawa Test&Measurement developed the AQ6380 with an easy-to-use front-panel design and intuitive operability.

Large LCD touchscreen

The high-resolution, responsive 10.4-inch multi-touch capacitive touchscreen makes device operation simple and intuitive. Easily change the view and measurement conditions and perform analysis with just a touch. In the optical spectrum view, the waveform view can be zoomed or shifted by a simple tap and drag.

LCD Touchscreen AQ6380 Optical Spectrum Analyzer | Yokogawa Test&Measurement

 

Enhanced function keys

Operability is improved with the addition of frequently-used function keys on the front panel. These include sweep control (Auto/Single/Repeat/Stop), resolution, and sensitivity. The function keys can also be displayed as a pop-up on the screen.

AQ6380 Optical Spectrum Analyzer Function Keys | Yokogawa Test&Measurement

 

WDM transmission systems

To test a multi-channel, narrowly-spaced WDM system, high-spectral measurement performance is required to test the system’s internal circuit boards (e.g., laser modules, optical transceivers, final output signal). This includes simultaneous peak and OSNR measurements. AQ6380’s wide close-in dynamic range allows accurate OSNR measurement of DWDM transmission systems. The built-in WDM analysis function analyzes the measured waveform and shows peak wavelength, peak level, and OSNR of WDM signals up to 1024 channels simultaneously.

AQ6380 Optical Spectrum Analyzer WDM Analysis | Yokogawa Test&Measurement
Example of DWDM OSNR analysis

 

Lasers/optical transceivers

Testing of optical components used in WDM transmission systems such as laser devices, laser modules, and optical transceivers requires high accuracy. Example applications include modulated signal measurement of optical transceivers and transponders and measurement of all channels of optical transceivers with WDM technology beyond 100 G. The AQ6380’s sharp spectral characteristics and high straylight suppression performance enables users to visualize and accurately measure spectral peaks in close proximity.

AQ6380 Optical Spectrum Analyzer Laser Optical Transceivers | Yokogawa Test&Measurement
Example of DFB-LD analysis

 

Passive components (FBG/WSS)

In conjunction with a broadband light sources such as ASE, SLD, or Super Continuum (SC), the AQ6380 enables higher resolution and wider dynamic range measurements and performs evaluation of passive devices including WDM, FBG, ROADM, and WSS. In addition, the built-in optical filter analysis function simultaneously reports peak/bottom wavelength, level, crosstalk, and ripple width.

USB ports

The AQ6380 has USB ports that are compatible with a USB storage device, mouse, and keyboard. The file feature allows users to save data and screenshots to internal memory or USB storage for use in creating test reports. By connecting a mouse or keyboard to the USB port, you can comfortably operate the AQ6380 as if you were operating a PC.

Data access through LAN

The standard LAN port allows convenient access to files stored in the internal memory and enables remote updates for firmware from a PC.

Functions

AQ6380 Optical Spectrum Analyzer Functions | Yokogawa Test&Measurement

Specifications

AQ6380 Optical Spectrum Analyzer Specifications | Yokogawa Test&Measurement

Note: All specifications are valid under the following conditions unless otherwise stated: 23˚C ±3˚C, after warming up for 1 hour or more, pulsed light measurement mode off, high dynamic range mode off, horizontal scale: wavelength, and vertical scale: absolute power.
*1: Using a 9.5/125 μm single mode fiber with a PC type connector, after optical alignment and wavelength calibration with built-in reference light source
*2: Wavelength resolution setting: ≥ 0.1 nm
*3: Resolution: 0.005 nm, number of sampling: AUTO, excluding sensitivity RAPID1 and NORM_HOLD
*4: After performing RBW calibration using a single longitudinal mode laser (wavelength: 1520 to 1560 nm, peak level: ≥ −20 dBm, level stability: ≤ 0.1 dBpp, and wavelength stability: ≤ ±0.001 nm), at the wavelength of RBW calibration
*5: TRAD mode: applicable for CW light and pulsed light measurement, RAPID mode: applicable for CW light measurement
*6: Resolution: 0.1 nm, number of sampling: AUTO
*7: 1523 nm, high dynamic range mode: SWITCH
*8: HeNe laser (1523 nm), wavelength resolution: 0.1 nm, 1520 nm to 1620 nm (except peak wavelength ±2 nm)
*9: Using Yokogawa’s master single mode fiber with a PC connector
*10: Typical value
*11: Some AQ6317 series commands may not be compatible due to changes in specifications or functions.
*12: Liquid crystal display may include a few defective pixels (within 0.002% with respect to the total number of pixels including RGB). There may be a few pixels on the liquid crystal display that do not emit all the time or remains ON all the time. These are not malfunctioning.
LCA  Reduced emission of CO2 about 74% compared to the previous model.

PDFResults of Life Cycle Assessment

AQ9441 Connector Adapter

Suffix Descriptions
-FC FC type
-SC SC type

AQ6370 Viewer

The AQ6370 Viewer is a package of PC application software for the AQ6380, the AQ6360, and the AQ6370 series Optical Spectrum Analyzer.

751535-E5 Rack mounting kit

RACK MOUNTING KIT For an EIA-compliant Single-housing Rack

Overview:

Optimizing bias voltage is essential for efficient optical modulator operation, maintenance of signal quality, and meeting performance specifications required for a designated application. Proper biasing helps achieve desired modulation effects, reduces distortion, and enhances overall reliability of optical communication systems.

Overview:

Optical transceivers are one of the indispensable key devices for optical communications that interconvert optical and electrical signals.

Application Note
Overview:

OSA Macro Program Function

The OSA Macro Program Function enables automated measurement by creating programs for entry of measurement conditions and other tasks. Users can program a sequence of measurement procedures from entry of measurement conditions (e.g., wavelength sweep width, setting resolution) to analyses, data saving, output, and others and eliminates redundant procedures on the production line. The function acts as a controller of other connected devices through the LAN or RS232C port and allows users to build an automatic measuring system without using an external PC and input measurement conditions or output measured results while programs are running.

Overview:

Overview of optical communications via optical fibers including: signal conversion, optical fiber benefits, techniques like wavelength division multiplexing (WDM) for increased capacity, key components like optical amplifiers and spectrum analyzers for maintaining transmission quality.

Overview:

Lack of reliable high-speed internet access in rural regions, due to complicated logistics and the considerable costs involved to extend land-based networks to these areas, has inspired a wave of next-generation applications that will provide greater accessibility and reliability. Making use of “space laser” networks, these revolutionary solutions can relay digital traffic via low Earth orbit (LEO) satellite systems to provide low-latency, high-speed broadband services to communities typically beyond the reach of standard wireless and fiber networks.

Overview:

To accurately measure pulsed light using an optical spectrum analyzer (OSA), it is necessary to understand the characteristics of the OSA and select the appropriate measurement method and settings.

Overview:

The extreme test requirements of our research called for an OSA with extended MIR spectrum bandwidth capabilities up to 5μm, but we couldn’t find one on the market capable of measuring optical inputs at these wavelengths. Yokogawa Test&Measurement rose to the challenge and developed a new OSA model for us that would. Not only do we now have an instrument that is practically custom-made for our needs, it provides repeatable, accurate, and trusted measurement outputs and is easy to learn and use. Their equipment and ability to create a new optical measurement solution has definitely increased the overall efficiency and productivity of our research team.

— Martin Bernier, PhD, P.Eng., Full Professor, Centre de Optique, Photonique, et Laser, Université Laval

Product Overviews

    Overview:

    Introducing the new Yokogawa Test&Measurement AQ6380 Optical Spectrum Analyzer. This new OSA includes many sought-after features including:

    • An unprecedented 5 pm wavelength resolution
    • ±5 pm wavelength accuracy
    • 1200 nm to 1650 nm wavelength range
    • 65 dB wide close-in dynamic range
    • 80 dB stray light suppression
    • Automated wavelength calibration
    • Gas purging
    • DUT-oriented interface and test apps
    • Backward-compatible remote interface
    • 10.4in intuitive touchscreen
    • Up to 20x faster measurement
    • Remote operation capabilities

How-tos

    Overview:

    We are going live on YouTube to answer your questions about the Yokogawa Test&Measurement AQ6380 high performance optical spectrum analyzer. Join us to discuss how to make the most of this award-winning instrument. During this live stream we will review the resolution and dynamic range of the AQ6380 OSA along with an example application of a 1310nm signal from an AQ1000 XFP module. Whether you’ve worked with an OSA for years or curious if it is a good fit for your work or research, this live stream can help.

    Potential items for review include but are not limited to:

    • Span
    • Sweep
    • Calibration
    • Analysis
    • Cursors/Markers
    • PC-Based Remote Viewing Software
    Overview:

    Application Engineer Danielle walks us through how to use an optical spectrum analyzer (OSA) to measure a gas cell in just a few easy steps.

Webinars

    Overview:

    Increasing demands for greater network bandwidth and expansion of 5G wireless backhaul has driven data rates to 400G, 800G, and beyond. Starting with the physical layer of the optical network stack, precision measurement of these light speed signals is crucial.

    This webinar covers a range of alternatives used to evaluate test parameters for DWDM, OSNR, SMSR, and EDFA, and includes a discussion on interesting applications like the use of high-energy lasers across space satellite mesh networks.

    Key topics include:

    • Key test methods and technologies
    • Traditional and new applications
    • Common testing oversights
    • Measuring against industry standards
    • Ensuring network element performance and reliability
    Overview:

    A vision of self-driving cars propels the research and development of automotive LiDAR, a vital hardware providing distance and velocity information of a vehicle’s surroundings. Some LiDAR concepts are already heading toward production for automotive ADAS and industrial markets. Two newer concepts promise the greatest potential yet: frequency-modulated continuous wave (FMCW) LiDAR and time-of-flight (TOF) flash LiDAR. However, there are engineering challenges impeding their full adoption. This webinar reviews operation principles and challenges of different LiDAR concepts, a brief discussion on the LiDAR market, and a review of critical optical components such as photodetectors and sources.

    Key takeaways include:

    • Fundamental test solution principles for challenges in demanding environments
    • Ensuring flawless and reliable operation of potentially life-critical components and systems over the entire life of the equipment
    • Techniques and results of signals from both LiDAR and facial recognition lasers from different smartphones
    • Experience how LiDAR manages (or mismanages) navigating metropolitan city streets in a truly driverless Level 4 Robotaxi
    Overview:

    Mastering the fundamentals of optical wavelength measurements and having a solid understanding of measurement principles for optical sources and devices is key to measuring with confidence. This webinar provides a thorough review of these foundational elements and concepts as well as:

    • Fiber identification and recommendations for routine care
    • Measurement techniques for different optical measurement devices
    • Example wavelength-specific applications for visible light to over 3000 nm such as telecommunications, biomedical, and atmospheric gas sensing
    • Important considerations for selecting an optical spectrum analyzer
    Overview:

    There are countless technologies available for optical communications devices and systems validation. With so many specifics to take into consideration, it's not always easy for an engineer to determine the best networking and fiber optic measurement solution to address their measurement needs.

    Key discussions in this on-demand webinar include:

    • The what, why, and how of available options like optical spectrum analyzers, optical wavelength meters, optical power meters, variable attenuators, fixed and tunable laser sources, and more
    • How to improve the quality and value of results for both active and passive optical devices
    • Ways to streamline productivity and reduce costs while also achieving higher data transmission rates, longer-distance transmissions, immunity to EMI, lower signal loss, lower latency, enhanced security, and improved energy efficiency
    • Trends driven by applications such as AI, quantum, and inter-satellite laser communications (i.e., space lasers!)

Looking for more information on our people, technology and solutions?


Contact Us

Precision Making

Top