How to Set a Strain Gage Factor Greater Than 2.20

For a strain gage factor greater than 2.20, please set the strain gage factor to 2.0 and use linear scaling. Refer to the example below.
 
For X > 2.20:
    - set gage factor of strain module to 2.00
    - set linear scaling constant A equal to a value of 2.0/X.
       For example, if X equals to 2.21, set scaling constant A to 2.00/2.21 or 0.905.
 
Related Information:
This method is also valid for the WE7245 module of the WE7000 series, for a gage factor greater than 2.30.

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

ScopeCorders

  • Flexible, high-performance Yokogawa ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

 

Precision Making

Top