Is There a Method to Perform Deskew on a DL850 When Using a Current Probe?

It is not necessary to perform deskew on a DL850 when using a current probe. The DL850 does have enough resolution to perform deskew and will not be able to find the time offset. Typically the deskew value ranges from ± 100ns in 0.01 ns steps. This means that for skew levels less than 100ns, the resolution of the skew must be 0.01ns. However, the resolution for the DL850 with the max 100Ms/s module only has 10ns resolution. For the 10Ms/s module the resolution is 100ns. In both modules, there is insufficient resolution to detect the skew. Please download the file for an explanation of deskew, taken from the DLM2000 instruction manual IM710105-01E.

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

World-class digital oscilloscopes from Yokogawa: The digital oscilloscopes have high-speed sampling and a wide range of bandwidths that can be utilized for design and development of electronic devices. The ScopeCorders have the advantages of both a digital oscilloscope and a multi-channel data recorder.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.

Top