What is the Delay Time of the External Trigger Input on the DL2700?

There is zero delay because the external trigger circuit on the DL2700 is located before the A/D converter. The timing is approximately 10 ns faster than the trigger captured on the measured waveform, since there are no attenuators or other circuits.

Related Products & Solutions

Oscilloscopes

World-class digital oscilloscopes from Yokogawa: The digital oscilloscopes have high-speed sampling and a wide range of bandwidths that can be utilized for design and development of electronic devices. The ScopeCorders have the advantages of both a digital oscilloscope and a multi-channel data recorder.