Why Does the Pulse Width of the NO-GO Output for the DL1540 Digital Oscilloscope Occasionally Decrease to 12 ms and What Causes This Change?

The NO-GO output is controlled by software. Thus, there may be fluctuations in the processing time depending on the sensing conditions. Since the signal remains low until the next determination can be accepted, this means that the processing time may fluctuate at the software level depending on the conditions. In any case, the minimum pulse width of 10 ms is guaranteed.

Related Products & Solutions

Mixed Signal Oscilloscopes

  • Yokogawa mixed signal oscilloscopes
  • Simultaneous, time correlated observations
  • Analysis of analog with digital (logic) signals
  • Troubleshooting electrical anomalies, measuring parametric values, monitoring signals

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

Oscilloscopes Application Software

  • Yokogawa Test&Measurement oscilloscope application software
  • View, export, analyze, and manage data 

Precision Making

Top