How to trigger on a CAN ID using DL850 ScopeCorder?

How to trigger on a CAN ID using DL850 ScopeCorder
 
A standard DL850 cannot provide a CAN trigger … but the feature added is “CAN ID” a new operator added in Realtime.
 
The CAN ID G3 operator uses an ANALOG input as the SOURCE.
The input is the packet.
The output is a pulse.
And you can trigger on that pulse as SOURCE = RMATH1 (as an example) … in realtime. 
 
So – the trigger will line-up with the CAN ID … 
And, … this does require G3 and DL850V both; 
 
And, you must update to the very latest firmware, this was a new feature released in August 2012;  

 

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Precision Making

Top