How to use SNTP on DL850 ScopeCorder
The sequence below is very important for SNTP to work properly.
1. Set DATE/TIME and GMT Time Difference to your location FIRST. EDT is -4 hours, EDT is -5 hours - as examples.
2. Set a SNTP Server for GMT in SNTP settings (64.90.182.55 in New York City for example - this is GMT, not EDT or EST);
3. Set up TCP IP as usual.
4. Under SNTP function, press ADJUST .. this must be performed as the last step.
200, 350, and 500MHz mixed-signal oscilloscopes for every engineer. Best-in-class performance in usability, acquisition, analysis, and display—all at a price you can digest. Options include serial bus, vehicle bus, and power supply analysis functions.
Scalable DAQ systems with industry-leading isolation, noise immunity, built-in conditioning, and real-time analysis, ensuring accurate, reliable measurements and faster decisions.
Accelerate debugging and gain deeper insight with high-resolution oscilloscopes designed for speed, clarity, and precision.