What is the ScopeCorder's Accuracy Specifications in Terms of DMM Digits?

In terms of display resolution, the 720210 and 701250 modules (12-bit) are equivalent to a 4 digit DMM. The 701262 module (16-bit) is equivalent to a 5 digit DMM. However, in terms of DC accuracy, the 720210 and 701250 modules are equivalent to a 3 digit DMM, and the 701262 module is equivalent to a 4 digit DMM.

See the attached screen shot for an example of the DMM style monitor function display mode.

For example:
- The 720210’s DC accuracy is 0.5% of 10 div
- A CH1 full scale of +/-50.00mV means 100mV span
- 0.5% of 100mV is 0.5mV
- But 720210’s display digit is 50.00mV
- Therefore, the DC accuracy of 720210 is equivalent to 3 digits.
 

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Precision Making

Top