What is the ScopeCorder's Accuracy Specifications in Terms of DMM Digits?

In terms of display resolution, the 720210 and 701250 modules (12-bit) are equivalent to a 4 digit DMM. The 701262 module (16-bit) is equivalent to a 5 digit DMM. However, in terms of DC accuracy, the 720210 and 701250 modules are equivalent to a 3 digit DMM, and the 701262 module is equivalent to a 4 digit DMM.

See the attached screen shot for an example of the DMM style monitor function display mode.

For example:
- The 720210’s DC accuracy is 0.5% of 10 div
- A CH1 full scale of +/-50.00mV means 100mV span
- 0.5% of 100mV is 0.5mV
- But 720210’s display digit is 50.00mV
- Therefore, the DC accuracy of 720210 is equivalent to 3 digits.

Related Products & Solutions

Data Acquisition (DAQ)

Data acquisition (DAQ) is the measurement, recording, analyzing, and presentation of real world phenomena. It includes electrical measurements such as voltage, current, power as well as measurements through sensors and transducers including temperature, pressure, strain, vibration, and more.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

Capture, analyze, and troubleshoot electrical or real-world physical signals. Oscilloscopes observe the change of electrical signals over time, continuously graphed on a display as voltage or amplitude vs. time.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.

 
Top