Leakage current measurement with current range from 20mA to 200A
Light weight and compact design
Manual range switching and auto power-off functions
24mm in diameter of measurable conductor
Mean value display
Auto power-off
Manual range switching
Approved for conformity to safety standards EN61010-1, EN61010-2-032(CAT. III 300 V)
Always observe instruction manual cautionary notes.
Follow maximum voltage level restrictions (up to 300V AC).
Keep transformer jaws fully closed during current measurement.
AC Current Measurement:
Set range switch.
Press jaw open/close lever and close jaws over one conductor only. The measured current value is shown on display.
Description | Leakage Current Mini Clamp-on |
---|---|
Method of detection | Mean Value |
Display | LCD (Digital display: 1999 counts) |
Response Time | Approx. 2 seconds |
Range Switching | Manual-range |
Data hold | On all range |
Operating temperature and humidity | 0 - 40°C, 85% RH or less (no condensation) |
Influence of external magnetic field | 10 mA or less in proximity to a 14.4 mm-dia conductor carrying 100A |
Influence of conductor position | Within 5 dgt for 0 to 50A, or 2% for 50 to 200A (10 mm-dia conductor at inside the jaw) |
Influence of Residual Current | 10 mA or less in proximity to a 10 mm-dia conductor carrying 50A |
Safety Standard | Conforms EN61010-1, EN61010-2-032 |
Circuit voltage | 300 Vrms or less |
Withstanding voltage | 3.7 kV AC for one minute |
Power Supply | LR-44 x 2(3V) SR-44 x 2 |
Battery Life | Approx. 15 hours (continuous) |
Consumed current | Approx. 5 mA |
Auto power-off | Approx. 10 minutes |
Diameter of measurable conductor | 24 mm at maximum |
Dimensions | Approx. 60 (W) x 149 (H) x 26 (D) mm |
Weight | Approx. 120 g |
Accessories | Batteries, carrying case (93033) User's manual (IM CL320) |
DC Current Range | - |
---|---|
DC Current Resolution | - |
AC Current Range | 20/200mA,200A |
AC Current Resolution | 0.01mA |
AC Current Frequency characteristics | 40~400Hz |
AC Current Method of detection | Mean Value |
AC Voltage Measurement | - |
DC Voltage Measurement | - |
Continuity Check Measurement | - |
Frequency Measurement | - |
Temperature Measurement | - |
Recorder output Function | - |
Waveform monitor output Function | - |
Data hold Function | O |
Auto hold Function | - |
Peak Hold Function | - |
Maximum value memory Function | - |
Mean value display Function | - |
Filter Function | O |
Range Hold Function | O |
Auto ZERO Adjust Function | - |
Auto power-off Function | O |
At 23°C±5°C,75%RH or less
Accuracy: ±( % rdg + dgt) = ±(% readout + value of least significant digit)
Parameter | Reference | Accuracy | |
---|---|---|---|
WIDE(40~400Hz) | 50/60Hz | ||
AC current | 20mA/200mA | 2.0+4 (50/60Hz) | 3.0+5 (50/60Hz) |
200A(0~100A) | 5.0+6 (40~400Hz) | ||
200A(100.1~200A) | 5.0+4 (50/60Hz) | 5.0+5 (50/60Hz) |