Why Does the Self Test Fail on the WE7111 Digital Oscilloscope Module?

You must use a probe with a attenuator of "10:1", otherwise the self test will not run correctly. Also, before executing the self-test, please allow for a 30 minutes warm-up time. If the instrument does not warm up, the self test may not be accurate.

Related Products & Solutions

Data Acquisition (DAQ)

Simplify complex testing with DAQ systems that flex to your needs. Capture and analyze any physical or electrical signal with ease, transforming signals into actionable insights that speed innovation and improve reliability.

High Speed Data Acquisition

  • PC-based, streaming, local or remote operation
  • 20+ modules, isolated and versatile inputs
  • Up to 200 MS/s or 640 ch

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.