How to trigger on a CAN ID using DL850 ScopeCorder?

How to trigger on a CAN ID using DL850 ScopeCorder
 
A standard DL850 cannot provide a CAN trigger … but the feature added is “CAN ID” a new operator added in Realtime.
 
The CAN ID G3 operator uses an ANALOG input as the SOURCE.
The input is the packet.
The output is a pulse.
And you can trigger on that pulse as SOURCE = RMATH1 (as an example) … in realtime. 
 
So – the trigger will line-up with the CAN ID … 
And, … this does require G3 and DL850V both; 
 
And, you must update to the very latest firmware, this was a new feature released in August 2012;  

 