Frequency to Voltage Conversion

Frequency to Voltage Conversion
 
Problem:
A transducer produces a sine-wave output dependant on the voltage-input; here is how to decode that output with a scope or a ScopeCorder. 
For 3kV input, the transducer outputs 100Hz.
For 30kV, the transducer outputs 1,000Hz.
 
Input waveform goes into ScopeCorder function (or scope) with /G2.
Sine wave goes into CH1.
M1=BIN(CH1) ... M1 is square wave with same frequency as trandsucer output.
 
M2=PWHH(M1) ... here we are measuring the pulse width of the M1 waveform.
 
M3 = 30*1/M2  ..
 
Go to MEASURE menu and take MAX(M3) .. or AVG(M3) and you now have a numerical result on-screen. Please do not use ROLL MODE. 
 
If ROLL MODE is a requirement, or for improved accuracy, use a FREQUENCY Module (701280) with linear scaling AX+B where A=30 and B=0.

Related Products & Solutions

DL9000 DSO Series

500MHz, 1.0GHz, and 1.5GHz DSOs for debug and high performance applications. 10th
generation oscilloscope from Yokogawa with industry leading 2.5 million wfms/sec and lowest dead time. Winner of Test & Measurement World's "Best in Test" award.

DLM2000 Mixed Signal Oscilloscopes

200, 350, and 500MHz mixed-signal oscilloscopes for every engineer. Best-in-class performance in usability, acquisition, analysis, and display—all at a price you can digest. Options include serial bus, vehicle bus, and power supply analysis functions.

DLM4000 Series Mixed Signal Oscilloscope

8 channel mixed-signal oscilloscopes for every engineer in bandwidths of 350 MHz or 500 MHz. Best-in-class performance in usability, acquisition, analysis, and display.

DLM5000 Series Mixed Signal Oscilloscope

  • 4-to-8 channel Yokogawa DLM5000- best mixed signal oscilloscope
  • Responsive touchscreen for easy navigation through analysis features 

DLM5000HD Series High-Definition Oscilloscope

The DLM5000HD High-Definition Oscilloscope from Yokogawa Test&Measurement is the new standard of high-definition oscilloscope.

  • Frequency Bandwidth: < 500 MHz
  • Sample Rate: < 2.5 GS/s
  • Vertical Axis Resolution: 12 bit
  • High Resolution Mode: < 16 bit
  • Measurement Memory: < 1 G points
  • History Waveforms: < 200,000
  • Simultaneous Multi-Channel Measurements: < 32 bit logic + 8 channels
    • < 16 channels with two connected DLM5000HDs via DLMsync
  • Highly-Responsive, Intuitive 12.1in Touchscreen
    • with Traditional Oscilloscope Control Panel
  • IEEE1588 Synchronous Support
  • Superior Noise Immunity
  • Vehicle Serial Bus
  • Additional Math Channels
  • Extensive Voltage Ranges
  • Compact, Light, Portable

DLM6000 MSO & DSO Series

500MHz, 1.0GHz, and 1.5GHz DSO and MSO models for debug, waveform characterization, bench top, or automated test applications. 4 channel models with 16 or 32 logic inputs. 12th generation oscilloscope with ergonomic physical and on-screen improvements.

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Mixed Signal Oscilloscopes

  • Yokogawa mixed signal oscilloscopes
  • Simultaneous, time correlated observations
  • Analysis of analog with digital (logic) signals
  • Troubleshooting electrical anomalies, measuring parametric values, monitoring signals

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

ScopeCorders

  • Flexible, high-performance Yokogawa ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

 

Precision Making

Top