My GS610 Displays a "Temperature Abnormality" Error, What Should I do?

The GS610 will display a "Temperature Abnormality" error when it detects an internal overheat condition. When the GS610 detects an overheat condition, it will:
  1. Forcibly turns the output (voltage or current) OFF
  2. Display "Temperature Abnormality" warning but there is no error number

If this happens, immediately turn OFF the power switch of the GS610. Provide adequate space around the GS610 or check and remove foreign objects caught in the cooling fan on the rear panel. Please wait a substantial amount of time before you turn ON the power switch. If the same warning appears, the GS610 may have malfunctioned and will need to be sent in for repairs.

Precision Making

返回顶部
WeChat QR Code
微信扫一扫
获得更多专业服务