How do I Measure Delay on the DL850?

The DL series instruments are capable of measuring the delay between traces or the delay between channels. The measured output value can be displayed in either time or degree.

To setup the DL850 Scopecorder to measure delay, follow the listed procedures:

  1. Press Measure » Measure Mode ON using the soft key
  2. Press the Measure Setup soft key and select the Source Trace
  3. Press the Delay Setup soft key to display the setup menu
  4. Use the Jog Shuttle/Arrow keys to navigate to Mode and select either Time or Degree
  5. Set the Polarity of the edge you want to detect to rising or falling
  6. Set which Edge Count you wish to use as the start of the measurement point
  7. Set the Reference to either Trace or Trigger
  8. Press ESC

Please download the attached PDF file for a detailed example of how to measure delay.

Note: For non-triggered ROLL Mode with a T/Div ≥ 100 ms/Div, you will need to press STOP to display the delay measurement.

Related Products & Solutions

DL7440/DL7480 Digital Oscilloscopes

8-channel 500 MHz oscilloscope with 16 logic inputs, for applications where four channels aren't enough.  Support for eight analog channels, 16 bits with 250MHz logic bandwidth, 16MP of recording memory, power analysis, serial bus analysis, and a broad selection of voltage and current probes.

DL9000 DSO Series

500MHz, 1.0GHz, and 1.5GHz DSOs for debug and high performance applications. 10th
generation oscilloscope from Yokogawa with industry leading 2.5 million wfms/sec and lowest dead time. Winner of Test & Measurement World's "Best in Test" award.

DLM2000 Mixed Signal Oscilloscopes

200, 350, and 500MHz mixed-signal oscilloscopes for every engineer. Best-in-class performance in usability, acquisition, analysis, and display—all at a price you can digest. Options include serial bus, vehicle bus, and power supply analysis functions.

DLM6000 MSO & DSO Series

500MHz, 1.0GHz, and 1.5GHz DSO and MSO models for debug, waveform characterization, bench top, or automated test applications. 4 channel models with 16 or 32 logic inputs. 12th generation oscilloscope with ergonomic physical and on-screen improvements.

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Precision Making

Top