Will a Reverse Polarity Connection of DC Power to the DL750 ScopeCorder Damage the Unit?

Making a reverse polarity connection of DC power to the DL750 ScopeCorder with the DC power supply option will not damage the unit. The DL750 ScopeCorder has a built-in protection circuit that will prevent damage to the unit in this situation. To recover from a reverse polarity connection, simply perform the following:
  1. Remove the reverse polarity connection from DL750 ScopeCorder
  2. Turn the DL750 ScopeCorder OFF for 1 minute
  3. Connect the DC power to the DL750 ScopeCorder with the correct polarity
  4. Power ON the DL750 ScopeCorder
  5. DL750 ScopeCorder will recover and run self test and calibration

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

ScopeCorders

  • Flexible, high-performance Yokogawa ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

 

Precision Making

Top