How do I Update the Voltage and Current Range for the PZ4000 and WT1600 Using Communication Commands?

To change the voltage range on element 1 to 30 V, send the "INPUT:VOLTAGE:RANGE:ELEMENT1 30V" command to the measuring instrument.
To change the current range on element 1 to the 1 A range, send the "INPUT:CURRENT:RANGE:ELEMENT1 1A" command to the measuring instrument.

After entering settings, confirm the settings by sending the following commands:

  • To query the voltage on element 1: "INPUT:VOLTAGE:RANGE:ELEMENT1?"
  • To query the current on element 1: "INPUT:CURRENT:RANGE:ELEMENT1?" 

To set/query the ranges on all elements at once, use the following commands:

  1. Voltage global setting
    "INPUT:VOLTAGE:RANGE 30V"
  2. Voltage global query
    "INPUT:VOLTAGE:RANGE?"
  3. Current global setting
    "INPUT:CURRENT:RANGE 1A"
  4. Current global query
    "INPUT:CURRENT:RANGE?

For information on the voltage and current values that can be set using these commands, please refer to:

  • PZ4000 Power Analyzer Communications Interface User's Manual
    Pages 4-48 and 4-52
  • WT1600 Digital Power Analyzer Communications Interface User's Manual
    Pages 4-49 and 4-53
 

Related Products & Solutions

PZ4000 Power Analyzer

The PZ4000 is designed for High Frequency Power Measurements as well as Transient Power Measurements. Its real-time acquisition system can perform cycle-by-cycle power measurements. One to Four input elements offer Basic Power Accuracy of 0.2% with 2 MHz Bandwidth and 5MS/sec digitizing rates.

WT1600 Digital Power Meter

The High-end WT1600 is designed for a wide range of applications, from energy-saving to large load applications. The WT1600 has a six element input capability to provide maximum flexibility, and Basic Power Accuracy of 0.1% with 1 MHz Bandwidth.

Power Analyzers and Power Meters

Yokogawa, the world's largest manufacturer of energy and power analyzers and meters, provides a broad choice of digital power analyzers to satisfy all requirements.

Precision Making

Top