DL850 ScopeCorder - How to maximize repeatability of Phase Measurements

DL850 ScopeCorder - How to maximize repeatability of Phase Measurements
 
  1. Use a direct-input if possible; this will remove the delay due to passive-probes 700929 being different between a pair of input channels (in phase response);
  2. Be sure INPUT COUPLING is set properly; it is easy to overlook; so, if you do use an input probe, be sure to set to 10:1, as an example.
  3. Set your V/DIV or SPAN up so as to maximize VERTICAL RESOLUTION - us as near to 20 vertical divisions as possible; this will decrease the influence of noise on the zero-crossing measurement of each waveform.
  4. Optional: using a BW-limiting filter at the front-end would further improve stability; it may actually (in theory) introduce some detectable phase-shift, so be sure to use the same filter on both channels if filtering is desirable.

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

ScopeCorders

  • Flexible, high-performance Yokogawa Test&Measurement ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

Precision Making

Top