If you want to know the condition of the "Check Range" LED indicator for voltage and current, you will need to read three bits in the Extended Event Register. "Bit-6", "Bit-7", and "Bit-8" needs to be read to know the status of the "Check Range" indicator on the front panel for Element 1.
There are two possible cases that cause the "Check Range" indicator to illuminate red. The "Check Range" LED will illuminate red if either condition occurs. Keep in mind that the voltage and current ranges are defined in RMS values.
Bit-6 of the condition register is defined as the "measured data over" bit. This bit corresponds to Case 1 and will return an ASCII value of "1" if the RMS measured/computed data of the input element exceeds the RMS range value.
Bit-7 of the condition register is defined as the "voltage peak over" bit. This bit corresponds to Case 2 and will return an ASCII value of "1" if the voltage value of the input element exceeds the PEAK value.
Bit-8 of the condition register is defined as the "current peak over" bit. This bit corresponds to Case 2 and will return an ASCII value of "1" if the current value of the input element exceeds the PEAK value.
Note:
Measure characteristics of devices that generate, transform or consume electricity. Also called power meters or wattmeters, these devices measure parameters such as true power (watts), power factor, harmonics, and efficiency.
Yokogawa's Power Analyzer software manages numeric, waveform, and harmonic data measurements. It enables data logging and instrument configuration from your computer.