Sample code for pulling just motor readings through command prompts on WT1800E

Here I have set up a sample code to make a list with 5 values in it.

I chose the values from the motor evaluation function list. 

 

:numeric:normal:number 5        

Here I made the number of values in the list 5

:NUMERIC:NORMAL:ITEM1 SPEed

 Here I made the first item speed 

:NUMERIC:NORMAL:ITEM2 Torq

 Here I made the second item torq

:NUMERIC:NORMAL:ITEM3 sync

 “”

:NUMERIC:NORMAL:ITEM4 Slip 

“”

:NUMERIC:NORMAL:ITEM5 PM

 “”

:numeric:normal? 

Here I returned the item setting I just set to be sure they were right 

:numeric:normal:value?

 Here I queried the values of those items in my list

 

Here are the corresponding responses:

 

<--- :NUMERIC:NORMAL:ITEM3 Torq ( 

<--- :NUMERIC:NORMAL:ITEM1 SPEed ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM2 Torq ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM3 sync ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM4 SLip ( 30 ms )

<--- :NUMERIC:NORMAL:ITEM5 PM ( 1 ms )

<--- :numeric:normal? ( 1 ms )

---> :NUM:NUM 5;ITEM1 SPE;ITEM2 TORQ;ITEM3 SYNC;ITEM4 SLIP;ITEM5 PM ( 2 ms )

<--- :numeric:normal:value? ( 1 ms )

---> -0.001E+00,-0.002E+00,2.5496E+03,100.000E+00,0.000E+00 ( 2 ms )