Why won't the WT3000E High Accuracy Power Analyzer connect with GA10?

The WT3000E High Accuracy Power Analyzer will not connect with GA10 if the Header and Verbose are set to ON.

To turn Verbose OFF, download the DL-term software from here: <https://tmi.yokogawa.com/library/documents-downloads/software/dl_term/>.

Then, send the command “COMMunicate:VERBose OFF”.