What is the Response Time of the 701260 Analog Voltage Module When in RMS Observation Mode?

The measured output response time of the 701260 Analog Voltage Input Module, when observing RMS is:
  • Rising (0 to 90% of 10 div): 100ms
  • Falling (100 to 10% of 10 div): 250ms
An attached JPEG displays the results of the 701260 output response to a step-input with a 60 Hz sine wave. Two zoom windows are displayed, the bottom-left zoom windows is the step-down response and the right window is the step-up response. The response time can be measured only when observing RMS.  