How to do a better RESET with AQ6370C

There is a LabVIEW VI (function) in your libraries called YKAQ6370 Preset Factory Defaults.vi. This function does not appear to clear everything out, for example, it does not clear-out the Stored Programs. From the OSA User’s Manual, you have a discussion in Section 9.2 that states: You can restore all settings to their factory defaults.

The following four types of initialization are available depending on the items to be initialized.

• PARAMETER CLEAR The parameter setting value of each function is initialized. Waveform data of TRACE A - G is also initialized. Use this to return the settings of the instrument to the already-known states.

• PARAM&DATA CLEAR Parameter setting values as well as data including MEMORY and PROGRAM are initialized.

• CAL DATA CLEAR Alignment adjustment values and wavelength calibration values are initialized.

• ALL CLEAR Current parameter setting values and data as well as alignment adjustment values and wavelength calibration values are initialized.

But, I would prefer to clear any or all of these settings from my Application; can you tell me what command needs to be called to perform this operation?

Answer: Preset should work. Its a fairly new feature and actually a button added to the front panel. PARAMETER INITIALIZE ALL CLEAR :SYSTem:PRESet

Related Products & Solutions

AQ6370C Optical Spectrum Analyzer

Popular TELECOM wavelength Range of 600nm to1700nm makes this an ideal model for Telecommunications applications for both singlemode and multi-mode optics.
New for 2010 includes both High Performance and Standard Performance versions.

Optical Spectrum Analyzer

  • Optical Spectrum Analyzer is used to measure and display power distribution of an optical source
  • Optical analyzer traces and displays power in vertical scale and wavelength in horizontal scale

Precision Making

Top