Real-time math encoder count issue

The DL850E real-time math ABZ encoder option was limited to a pulse/ rotation count of 65535. Now with the latest firmware the limit has been adjusted to 500000. In the firmware update process, the same update file should update the instruments G5 option as well. If the G5 option is not higher then 4.00 it will not count up to 360 degrees but will stop at a random degree count.