Why does the “ACTION ON TRIGGER” configuration not stay ON when power is cycled, and it has to be configured again or the setup loaded?
It does clear with power-cycle. To fix this, press: UTILITY, PREFERENCE, and then POWER ON ACTION. Set it for ACTION to ON and set it to START if you wish for it to also recover from a power-outage and it will press the START button for itself, so to speak.
Combining the capabilities of an Oscilloscope and the characteristics of a Data Acquisition Recorder, the Yokogawa Scopecorder is a versatile tool for engineers making measurements in the industrial, transportation and energy sectors.
An all-in-one electromechanical measurement system offering the deepest measurement and analysis features to turn data into insights.