ACTION ON TRIGGER disables itself after POWER CYCLE on ScopeCorder - why?

Why does the “ACTION ON TRIGGER” configuration not stay ON when power is cycled, and it has to be configured again or the setup loaded?

It does clear with power-cycle. To fix this, press: UTILITY, PREFERENCE, and then POWER ON ACTION. Set it for ACTION to ON and set it to START if you wish for it to also recover from a power-outage and it will press the START button for itself, so to speak.